Out in Front: June 6, 1996
Thanks to a multiprocessor architecture called MegaZoom, the two DSOs avoid the noticeable delay that typically occurs between changes in front-panel settings and the resulting changes in the display. A similar lag occurs when the signal changes. With the HP scopes, the delay is imperceptible. As with HP's earlier 54620 units, which perform only logic-timing analysis, the 54645D emphasizes the requirements of people who don't use a logic analyzer every day. You select trigger conditions by using a straightforward menu system. And, you can use logic patterns to trigger analog acquisitions.
Despite their deep memories, the scopes incorporate random equivalent-time sampling for repetitive waveforms. The scopes can use both single-shot and equivalent-time sampling. When the signals are repetitive, random sampling makes the effective sampling rate several times as high as the maximum 200M-sample/sec real-time rate. This feature provides better waveform detail than is possible when the scope samples rapidly changing signals at the maximum single-shot rate.
by Dan Strassberg
Hewlett-Packard Co, Santa Clara, CA. (800) 452-4844, http://www.tmo.hp.com.